Description
The StarLighter Fluorometer X Series is a family of compact microvolume fluorescence instruments designed for rapid nucleic acid and protein quantification. Three configurations are available to meet different laboratory throughput and assay requirements: the single-sample StarLighter X1, the eight-sample DNA-focused StarLighter X2, and the multi-application eight-sample StarLighter X3.
Only 1–20 µL of sample is required for routine quantification workflows. A measurement can be completed in no more than five seconds per sample, helping laboratories conserve valuable samples while reducing the time required for concentration analysis.
The three models differ in both throughput and supported sample types. The X1 supports DNA and RNA measurement for one sample at a time. The X2 processes up to eight samples per run and is intended exclusively for DNA measurement. The X3 processes up to eight samples per run and supports DNA, RNA, ssDNA and protein measurement.
The X3 also provides dedicated Fluorometer and Kinetics programs for direct fluorescence measurement and fluorescence kinetics experiments. Its touchscreen interface supports assay selection, two-point calibration, sample-volume settings, result display, historical-data review and USB data export.
Choose the Appropriate StarLighter Model
| Model | Catalog Number | Throughput | DNA | RNA | ssDNA | Protein |
|---|---|---|---|---|---|---|
| StarLighter X1 | FS-M2001 | 1 sample per measurement | Yes | Yes | No | No |
| StarLighter X2 | FS-M2002 | Up to 8 samples per run | Yes | No | No | No |
| StarLighter X3 | FS-M2003 | Up to 8 samples per run | Yes | Yes | Yes | Yes |
Microvolume Quantification
The StarLighter platform requires only 1–20 µL of sample. This makes the instruments suitable for workflows involving limited sample availability or low-concentration nucleic acid and protein preparations.
Fast Laboratory Workflow
Each sample can be measured in no more than five seconds. The touchscreen-guided workflow covers assay selection, calibration, sample-volume entry, measurement and result review.
High-Sensitivity Fluorescence Detection
The instruments use an efficient monochromatic LED light source and a SiPMT fluorescence detector. The documented optical system uses blue excitation at 460–480 nm and green emission at 500–535 nm.
Eight-Sample Processing with X2 and X3
The X2 and X3 accept an eight-tube strip or individual 0.2 mL transparent high-profile PCR tubes. Operators can select the required positions and process as many as eight samples in one run.
Advanced X3 Measurement Programs
In addition to quantitative DNA, RNA, ssDNA and protein assays, the X3 provides direct Fluorometer and Kinetics modes. The Kinetics program can display a fluorescence curve together with starting fluorescence, ending fluorescence, maximum fluorescence, minimum fluorescence and the time required to reach recorded fluorescence points.
Technical Specifications
| Parameter | Specification |
|---|---|
| Product series | StarLighter Fluorometer X Series |
| Available models | X1 / FS-M2001; X2 / FS-M2002; X3 / FS-M2003 |
| Instrument type | Microvolume fluorometer |
| Sample volume | 1–20 µL |
| Measurement time | ≤5 seconds per sample |
| Linear range | Five orders of magnitude |
| Light source | Efficient monochromatic LED |
| Excitation channel | Blue: 460–480 nm |
| Emission channel | Green: 500–535 nm |
| Detector | SiPMT |
| Calibration | Two-point calibration |
| Power requirement | DC 5 V, 2 A |
| Maximum power | 10 W |
| Operating temperature | 15–30°C |
| Relative humidity | ≤75% |
| Operating environment | Indoor use only |
| Maximum operating altitude | Up to 2,000 m |
| Transportation and storage temperature | -20–60°C |
| Dimensions | 110 × 230 × 45.5 mm |
| Weight | 0.5 kg |
| Warranty | One-year instrument warranty with free replacement for qualifying warranty cases during the warranty period |
Model-Specific Configuration
| Parameter | X1 / FS-M2001 | X2 / FS-M2002 | X3 / FS-M2003 |
|---|---|---|---|
| Sample throughput | 1 | Up to 8 | Up to 8 |
| Supported measurements | DNA and RNA | DNA only | DNA, RNA, ssDNA and protein |
| Sample-tube format | 0.5 mL transparent thin-wall PCR tube | Eight-tube strip or individual PCR tubes | 0.2 mL transparent high-profile PCR eight-tube strip or individual tubes |
| Additional modes | Model-specific assay interface | Model-specific assay interface | Fluorometer and Kinetics |
| Communication interface | USB data functions | Model-specific configuration | Bluetooth and serial interface; USB data export |
X1 Documented Detection Ranges
| Analyte | Detection Range |
|---|---|
| dsDNA | 0.01–1000 ng/µL |
| RNA | 0.25–1000 ng/µL |
The X1 configuration offered on this product page is intended for DNA and RNA measurement only. It is not offered for ssDNA or protein measurement.
X3 Quantification Programs and Ranges
| Program | Sample Concentration Range | Sample Amount |
|---|---|---|
| dsDNA High Sensitivity | 0.005–120 ng/µL | 0.1–120 ng |
| dsDNA Broad Range | 0.2–2000 ng/µL | 2–2000 ng |
| microRNA | 0.05–100 ng/µL | 1–1000 ng |
| ssDNA / Oligo | 0.05–200 ng/µL | 1–200 ng |
| Protein BR Assay Kit | 100–2000 ng/µL | 1–400 µg |
The X3 software identifies its RNA measurement program as “microRNA” and its single-stranded DNA program as “Oligo.”
Features
Three Configurations for Different Workloads
Select a single-sample DNA/RNA instrument, an eight-sample DNA instrument or an eight-sample multi-application instrument without changing to a different product family.
Low Sample Consumption
Only 1–20 µL of sample is required for routine quantitative measurements, helping preserve material for downstream analysis.
Results in Five Seconds or Less
The stated measurement time is no more than five seconds per sample after the prepared reaction tube is loaded.
Eight Independently Selectable Positions
On X2 and X3 models, operators can select the required tube positions and process up to eight samples in one run.
Two-Point Calibration
Quantitative assays use two-point calibration. On the X3, the operator can perform a new calibration using two standards or reuse the previous calibration result when appropriate for the assay.
Touchscreen-Guided Operation
The user interface guides the operator through assay selection, calibration, sample-volume settings, tube-position selection, measurement and result review.
On-Instrument Data Management
Saved experimental records can be searched and reviewed on the instrument. Selected X3 records can be exported after a USB flash drive is inserted.
Direct Fluorescence Measurement
The X3 Fluorometer mode can be used to read fluorescence results independently of its predefined quantitative assay programs.
Fluorescence Kinetics
The X3 Kinetics mode is designed for experiments that require a fluorescence curve over time, including reagent and pharmaceutical research workflows.
Compact Benchtop Footprint
The documented instrument dimensions are 110 × 230 × 45.5 mm, with a weight of approximately 0.5 kg.
Measurement Workflow
- Select the appropriate measurement program for the installed model.
- Use the previous valid calibration or perform a new two-point calibration with the required standards.
- Add 1–20 µL of sample to the compatible transparent PCR tube.
- Add 180–199 µL of working solution so that the combined reaction volume is 200 µL.
- Close the tube and gently vortex for 2–3 seconds while minimizing bubble formation.
- Incubate at room temperature and protect the prepared reaction from light for two minutes, unless the selected assay instructions specify a different incubation period.
- Load the tube or selected tube positions into the instrument.
- Enter the sample volume and start the measurement.
- Review the concentration or fluorescence result on the touchscreen.
- Open the stored record for later review or export supported records through a USB flash drive.
Protein assays may require a longer incubation period. Follow the instructions supplied with the selected protein assay kit.
Applications
StarLighter X1 Applications
- Routine DNA concentration measurement
- Routine RNA concentration measurement
- Single-sample microvolume laboratory workflows
- Low-concentration nucleic acid quantification
StarLighter X2 Applications
- DNA concentration measurement
- Processing up to eight DNA samples per run
- DNA extraction quality-control workflows
- DNA sample normalization before downstream analysis
StarLighter X3 Applications
- High-sensitivity dsDNA quantification
- Broad-range dsDNA quantification
- RNA and microRNA quantification
- ssDNA and oligonucleotide quantification
- Broad-range protein quantification
- Direct fluorescence measurement
- Fluorescence kinetics experiments
- Reagent research involving fluorescence kinetic curves
- Pharmaceutical research involving fluorescence kinetic curves
- Multi-sample nucleic acid and protein workflows
Operating and Maintenance Guidance
- Operate the instrument indoors at 15–30°C and relative humidity of no more than 75%.
- Place the instrument on a stable, level and dry surface.
- Keep the instrument away from direct sunlight, heat sources, liquids, corrosive gases and strong magnetic interference.
- Avoid introducing bubbles when preparing samples because bubbles may affect fluorescence readings.
- Remove reaction tubes after measurement rather than leaving them inside the instrument.
- Keep standards and prepared samples protected from light between measurements.
- Clean the instrument weekly to help maintain performance and reduce contamination.
- Disconnect the power before cleaning.
- Use a dry cloth or a cloth or cotton swab lightly moistened with water or alcohol.
- Do not rinse the instrument or allow liquid to enter the housing.
- Do not use strong solvents or aggressive cleaning agents.
- Allow the instrument to dry naturally before reconnecting the power.
Warranty
The StarLighter Fluorometer X1, X2 and X3 are supplied with a one-year instrument warranty. If the instrument develops a qualifying warranty issue during the one-year warranty period, it is eligible for free replacement.
The warranty does not cover damage caused by unauthorized disassembly, improper operation, liquid ingress, unsuitable installation conditions or use outside the stated environmental and electrical requirements.
Frequently Asked Questions
What is the difference between the StarLighter X1, X2 and X3?
The X1 measures one sample at a time and supports DNA and RNA. The X2 processes up to eight samples per run and supports DNA only. The X3 processes up to eight samples per run and supports DNA, RNA, ssDNA and protein.
Can all three models measure protein?
No. Protein measurement is supported by the X3 only. The X1 and X2 configurations offered on this page do not support protein measurement.
Can the X1 measure ssDNA?
No. The X1 configuration offered on this page supports DNA and RNA but does not include ssDNA measurement.
Does the X2 measure RNA?
No. The X2 is the eight-sample DNA-only model.
Which model supports ssDNA or oligonucleotide measurement?
The X3 supports ssDNA measurement through the software program identified as Oligo.
Which model should I choose for the highest throughput?
Choose the X2 or X3 when up to eight samples must be processed in one run. Select the X2 for DNA-only workflows or the X3 when RNA, ssDNA or protein measurement is also required.
How much sample is required?
The documented microvolume workflow requires 1–20 µL of sample.
How long does a measurement take?
The stated measurement time is no more than five seconds per sample.
What PCR tubes does the X3 use?
The X3 uses a 0.2 mL transparent high-profile PCR eight-tube strip or compatible individual transparent PCR tubes.
What quantitative programs are installed on the X3?
The X3 manual lists dsDNA High Sensitivity, dsDNA Broad Range, microRNA, ssDNA/Oligo and Protein Broad Range programs. It also provides direct Fluorometer and Kinetics modes.
Can measurement data be exported?
Yes. The X3 allows selected experimental records to be exported after a USB flash drive is inserted.
Why can repeated fluorescence readings decrease?
Repeated readings can decrease when a tube remains inside the instrument and becomes warmer. Continuous excitation can also reduce the fluorescence signal of the dye. Remove the tube after reading and keep it protected from light between measurements.
Does temperature affect the result?
Yes. Fluorescence intensity can change with temperature. Standards and samples should be allowed to reach consistent room-temperature conditions before measurement.
What is the warranty period?
The instrument warranty is one year. Qualifying warranty cases are eligible for free replacement during that period.
When can I expect my order to ship?
Most orders are filled and shipped within 2-3 business days from the time they are received.
Our standard shipping usually take 2-5 days.
We also provide express shippping for time-sensitive deliveries.
Email contact@biofargo.com if you have any requirements.