A sandwich-format solid-phase enzyme-linked immunosorbent assay (ELISA) for quantitative determination of natural or recombinant human Interleukin-6 (hIL-6) in biological samples. Wells are pre-coated with a mouse anti-hIL-6 capture antibody. Samples and calibration standards are incubated in the wells, followed by a biotinylated anti-hIL-6 detection antibody and streptavidin-HRP (SA-HRP). TMB substrate is converted by HRP to a blue product (max absorbance ~655 nm) and then stopped to yield a yellow product read at 450 nm (with 620–650 nm reference). hIL-6 concentrations are calculated by fitting a 4-parameter logistic (4-PL) calibration curve from supplied standards.
Technical Specifications
| Parameter | Details |
|---|---|
| Package size | 96 tests/Kit |
| Assay format | Double-antibody sandwich ELISA (solid-phase) |
| Sample volume per well | 100 µL (standards, controls and samples) |
| Linearity & Range | 2–512 pg/mL, R² ≥ 0.990 |
| Lower Limit of Quantification (LLOQ) | 2 pg/mL |
| Specificity / Cross-reactivity | No cross-reactivity with IL-2, IL-10, IL-12, IFN-γ, IFN-β, IL-6Rα, Rat IL-6 or Mouse IL-6 |
| Typical calibration curve model | 4-parameter logistic (4-PL) regression; example parameters A=7.86930, B=-1.04651, C=760.10625, D=0.00086, R²=0.99999 |
| Calibration standards (pg/mL) | 512, 256, 128, 32, 8, 2, 1, 0 (NCS) |
| Incubation times | Sample incubation 1 hour at RT (600 rpm recommended); Biotinylated antibody 45 minutes at RT; Streptavidin-HRP 30 minutes at RT; TMB substrate 10 minutes at RT |
| Washing | Wash plates with 300 µL 1×Wash Buffer; typical wash cycles: after sample incubation 3×, after biotin 3×, before TMB 5× (as described) |
| Detection / Readout | Optical density read at 450 nm, with reference/correction wavelength 620–650 nm; read within 10 minutes after stop |
| Reagent concentrations / dilutions | Anti-hIL-6: Biotinylated Conjugate (200×) — dilute 200× with Diluent prior to use; Streptavidin-HRP (100×) — dilute 100× with Diluent prior to use; Wash Buffer Concentrate (10×) — dilute 10× |
| Storage of key components (opened) | Anti-hIL-6 Microtiter Strips: store in bag with desiccant at 2–8 °C up to 90 days; Reconstituted Calibration Standard: aliquot and store ≤-18 °C; other components: follow label instructions |
Features
- High sensitivity: LLOQ 2 pg/mL and linear range down to 2 pg/mL enabling detection of low-level hIL-6.
- High specificity: No observed cross-reactivity with related cytokines and IL-6 from rat or mouse.
- Ready-to-use coated strips: Microtiter strips pre-coated with capture antibody in sealed bag with desiccant for convenience and reduced setup time.
- Complete kit with standards and reagents: Includes lyophilized calibration standard, reconstitution solution, diluent, conjugates, SA-HRP, substrate and stop solution for a full quantitative assay.
- Standardized quantitation: Supplied calibration series and 4-PL fitting enable robust quantification of hIL-6.
Applications
- Quantitation of human IL-6 in serum
- Quantitation of human IL-6 in cell culture supernatants
- In-process samples, harvested bulk, drug substance and drug product testing (research applications)
- Research use in immunology, cell biology and bioprocess monitoring
Kit Contents
Attention
• For Research Use Only (not for clinical or diagnostic use).
• Qualified personnel only; validate assay in-house before routine use.
• Do not mix reagents from different lot numbers.
• Reconstituted Calibration Standard, 1× Biotinylated Conjugate and 1× Streptavidin-HRP are recommended single-use; prepare freshly.
• Use fresh ultrapure/sterile water; if concentrates are cloudy, heat to 37 °C until clear.
• Avoid drying wells before substrate incubation and avoid bubbles during pipetting.
• Stop Solution is 1 M HCl — corrosive; avoid contact with eyes, skin and clothing.
• Avoid samples containing sodium azide (NaN3) as it will inactivate HRP and lead to underestimation.
• TMB Substrate should be colorless; discard if discolored.
• Use sterile disposable consumables and avoid cross-contamination; recommended to wipe benches and equipment with 75% ethanol.
Quality Management & Certifications
Quality System
Download QMSQMS (ISO, GMP)
Download CertificateQuality Advantages
View ReportQuality Control Process
Download ProcessTechnical Resources & Downloads
Frequently Asked Questions (FAQ)
Research Use Only
For Research Use Only (RUO) — Not for clinical or diagnostic use – This product is intended for laboratory research purposes only and not for clinical or regulatory diagnostic use.
Not intended for use in USDA or FDA regulated diagnostic testing or official compliance testing.
When can I expect my order to ship?
Most orders are filled and shipped within 2-3 business days from the time they are received.
Our standard shipping usually take 2-5 days.
We also provide express shippping for time-sensitive deliveries.
Email contact@biofargo.com if you have any requirements.