Product Description
This product is a one-step, solid-phase, double-antibody sandwich Enzyme-Linked Immunosorbent Assay (ELISA) intended to detect residual host cell proteins (HCPs) from E.coli protein expression strains (originated from BL21) in bioprocess samples. Polyclonal capture antibody (sheep anti-E.coli HCP affinity antibody) is pre-coated on microtiter strips. Calibration standards and samples are simultaneously incubated with HRP-conjugated anti-E.coli HCP detection antibody (one-step format). HRP catalyzes TMB oxidation to produce a colorimetric signal measured at 450 nm (with long-wavelength correction 620–650 nm). Concentrations are calculated by a 4-parameter logistic (4PL) fit to the calibration curve. The kit is supplied for research use only and is not intended for clinical diagnostic use.
Technical Specifications
| Parameter | Details |
|---|---|
| Format / Throughput | 96 tests per kit; 8-well × 12 strips |
| Detection principle | One-step double-antibody sandwich ELISA with HRP/TMB colorimetric detection (read at 450 nm, reference 620–650 nm) |
| Linearity & Range | 1.5–243 ng/mL, R2 > 0.990 |
| Lower Limit of Quantitation (LLOQ) | 1.5 ng/mL |
| Calibration curve model | 4-parameter logistic (4PL); example parameters A=2.41626, B=1.09106, C=89.16833, D=0.01118, R2=0.99995 (example data) |
| Specificity / Cross-reactivity | No cross-reactivity observed with CHO, Vero, HEK293T, and Hansenula polymorpha HCPs (as tested) |
| Assay incubation conditions | Add 100 µL HRP detection reagent + 100 µL sample/standard per well; incubate on microplate thermoshaker at 600 rpm for 3 hours at room temperature; wash; add 100 µL TMB and incubate 15 minutes at room temperature in dark; stop and read |
| Reading | Absorbance measured at 450 nm with correction wavelength set at 620–650 nm; subtract NCS (blank) and use 4PL fit |
| Recommended sample pH | pH 6.5–8.5 (measurements may be compromised outside this range) |
| Minimum recommended pipetting volume for dilutions | Minimum volume 5 µL recommended for accurate dilutions |
| Reading window after stop | Complete plate reading within 30 minutes after termination |
Features
- One-step ELISA format: Simultaneous addition of calibration standard/sample and HRP-conjugated detection antibody reduces hands-on time and simplifies workflow.
- High sensitivity: LLOQ of 1.5 ng/mL with a linear range up to 243 ng/mL.
- Strain-targeted specificity: Polyclonal antibodies generated against E.coli protein expression strains (BL21 origin) to improve detection of residual HCPs from these expression hosts.
- Pre-coated strips: Microtiter strips are pre-coated with capture antibody and supplied in sealed bag with desiccant for convenience and reproducibility.
- Validated lack of cross-reactivity: No cross-reactivity observed with CHO, Vero, HEK293T, or Hansenula polymorpha HCPs (as tested).
- Complete kit: Includes calibration standard, HRP-conjugated detection antibody, diluent, wash buffer concentrate, substrate, stop solution and sealing films.
Applications
- Quantitation of residual host cell proteins (HCPs) in products expressed in E.coli BL21 protein expression strains
- In-process samples, harvested bulk, drug substance and drug product testing during downstream processing
- Assay development and validation for bioprocess monitoring of E.coli-expressed biologics (e.g., IL, rhIFN, rhGM-CSF, rhTNF, EGF/FGF/PDGF)
- Spike/recovery and method validation studies for residual HCP measurement (research use)
Kit Contents
Attention
• For research use only (RUO); not intended for clinical diagnostic use.
• Use sterile disposable tips, tubes and reservoirs; recommended to wipe bench and equipment with 75% ethanol before and after use.
• Do not mix reagents from different lot numbers.
• Use fresh ultra‑pure water; water temperature should not exceed 37 °C when preparing wash buffer.
• Reconstituted Calibration Standard and prepared 1× Anti‑E.coli HCP‑E:HRP are recommended for single use due to instability; prepare freshly before each experiment.
• TMB Substrate should be colorless; discard and contact supplier if colored.
• Stop Solution is 1 M HCl — avoid contact with eyes, skin and clothing; wear appropriate PPE.
• Avoid samples containing sodium azide (NaN3) as it will deactivate HRP and lead to underestimation of HCP levels.
• Recommended sample pH 6.5–8.5; outside this range results may be compromised.
• Minimum recommended pipetting volume for dilutions: 5 µL to ensure accuracy.
• Seal microplate immediately after sample loading to prevent evaporation; avoid drying wells before substrate incubation.
• Complete plate reading within 30 minutes after stopping the reaction.
Quality Management & Certifications
Quality System
Download QMSQMS (ISO, GMP)
Download CertificateQuality Advantages
View ReportQuality Control Process
Download ProcessTechnical Resources & Downloads
Frequently Asked Questions (FAQ)
Research Use Only
Research Use Only (RUO) – This product is intended for laboratory research purposes only and not for clinical or regulatory diagnostic use.
Not intended for use in USDA or FDA regulated diagnostic testing or official compliance testing.